Introduction and Specifications
General Specifications 1
1-17
1-20. AC Current (Sine Wave) Specifications
Absolute Uncertainty, tcal ± 5 °C
± (% of output + µA)
Range Frequency
90 days 1 year
Resolution
Compliance
Voltage
Max
Inductive
Load
10 to 20 Hz 0.19 0.15 0.25 0.15
20 to 45 Hz 0.09 0.15 0.125 0.15
45 Hz to 1 kHz 0.09 0.25 0.125 0.25
1 to 5 kHz 0.30 0.15 0.4 0.15
0.029 to 0.32999 mA
5 to 10 kHz 0.94 0.15 1.25 0.15
0.01 µA
3.0 V rms
1 µH
10 to 20 Hz 0.15 0.3 0.2 0.3
20 to 45 Hz 0.08 0.3 0.1 0.3
45 Hz to 1 kHz 0.08 0.3 0.1 0.3
1 to 5 kHz 0.15 0.3 0.2 0.3
0.33 to 3.2999 mA
5 to 10 kHz 0.45 0.3 0.6 0.3
0.01 µA
3.0 V rms
1 µH
10 to 20 Hz 0.15 3 0.2 3
20 to 45 Hz 0.08 3 0.1 3
45 Hz to 1 kHz 0.07 3 0.09 3
200 µH,
10 to
500 Hz
1 to 5 kHz 0.15 3 0.2 3
3.3 to 32.999 mA
5 to 10 kHz 0.45 3 0.6 3
0.1 µA
3.0 V rms
1 µH,
500 Hz to
10 kHz
10 to 20 Hz 0.15 30 0.2 30
20 to 45 Hz 0.08 30 0.1 30
45 Hz to 1 kHz 0.07 30 0.09 30
200 µH,
10 to
500 Hz
1 to 5 kHz 0.15 30 0.2 30
33 to 329.99 mA
5 to 10 kHz 0.45 30 0.6 30
1 µA
3.0 to
2.0 V rms
[1]
5 µH,
500 Hz to
10 kHz
10 to 45 Hz 0.15 300 0.2 300
45 Hz to 1 kHz 0.08 300 0.1 300
200 µH,
45 to
500 Hz
0.33 to 2.19999 A
1 to 5 kHz 0.7 300 0.75 300
10 µA
3.0 to
2.0 V rms
[2]
5 µH,
500 Hz to
5 kHz
45 to 65 Hz 0.05 2000 0.06 2000
65 to 500 Hz 0.08 2000 0.10 2000
200 µH,
45 to 65 Hz
2.2 to 11 A
500 Hz to 1 kHz 0.25 2000 0.33 2000
100 µA
2.8 to
1.25 V rms
[3]
1 µH,
65 Hz to
1 kHz