5500A
Operator Manual
1-18
AC Current (Sine Wave) Specifications (cont.)
Absolute Uncertainty, tcal ± 5 °C
± (% of output + µA)
Range Frequency
90 days 1 year
Resolution
Compliance
Voltage
Max
Inductive
Load
5725A Amplifier
45 Hz to 1 kHz 0.08 100 0.1 100
1 to 5 kHz 0.19 5000 0.25 5000
1.5 to 11 A
5 to 10 kHz 0.75 10000 1 10000
100 3
400 µH
[1] The actual voltage compliance (Vc) is a function of current output (Io), and is given by the formula:
Vc = -3.37*Io+3.11. The highest compliance voltage is limited to 3.0 V.
[2] The actual voltage compliance (Vc) is a function of current output (Io), and is given by the formula:
Vc = -0.535*Io+3.18. The highest compliance voltage is limited to 3.0 V.
[3] The actual voltage compliance (Vc) is a function of current output (Io), and is given by the formula:
Vc = -0.176*Io+3.19. The highest compliance voltage is limited to 2.8 V.
Range
Frequency
Maximum Distortion and Noise
10 Hz to 100 kHz Bandwidth
± (% output + µA)
10 to 20 Hz 0.15 1.0
20 to 45 Hz 0.1 1.0
45 Hz to 1 kHz 0.05 1.0
1 to 5 kHz 0.5 1.0
0.02 to 0.32999 mA
5 to 10 kHz 1.0 1.0
10 to 20 Hz 0.15 1.5
20 to 45 Hz 0.06 1.5
45 Hz to 1 kHz 0.02 1.5
1 to 5 kHz 0.5 1.5
0.33 to 3.2999 mA
5 to 10 kHz 1.2 1.5
10 to 20 Hz 0.15 5
20 to 45 Hz 0.05 5
45 Hz to 1 kHz 0.07 5
1 to 5 kHz 0.3 5
3.3 to 32.999 mA
5 to 10 kHz 0.7 5
10 to 20 Hz 0.15 50
20 to 45 Hz 0.05 50
45 Hz to 1 kHz 0.07 50
1 to 5 kHz 0.2 50
33 to 329.99 mA
5 to 10 kHz 0.4 50
10 to 45 Hz 0.2 500
45 Hz to 1 kHz 0.1 500
0.33 to 2.19999 A
1 to 5 kHz 1.4 500
45 to 65 Hz 0.2 3 mA
65 to 500 Hz 0.1 3 mA
2.2 to 11 A
500 Hz to 1 kHz 0.4 3 mA
5725A Amplifier
45 Hz to 1 kHz 0.05 1 mA
1 to 5 kHz 0.12 1 mA
1.5 to 11 A
5 to 10 kHz 0.5 1 mA