5500A
Operator Manual
7-10
7-10. Resistance Accuracy
The Resistance Accuracy test verifies the accuracy of synthesized resistance at the
5500A Calibrator front panel NORMAL terminals. For resistances of less than 110 k,
use the 4-wire COMP option. For resistances of 110 k or higher, the COMP option is
automatically turned off.
Nominal Value () Measured Value () Deviation % 90-Day Spec. (m or %)
0 6 m
2 0.309%
10.9 0.064%
11.9 0.135%
19 0.088%
30 0.059%
33 0.052%
109 0.021%
119 0.020%
190 0.015%
300 0.012%
330 0.025%
1.09 k 0.012%
1.19 k 0.012%
1.9 k 0.010%
3 k 0.009%
3.3 k 0.025%
10.9 k 0.012%
11.9 k 0.012%
19 k 0.010%
30 k 0.009%
33 k 0.026%
109 k 0.013%
119 k 0.014%
190 k 0.012%
300 k 0.011%
330 k 0.028%
1.09 M 0.016%
1.19 M 0.016%
1.9 M 0.014%
3 M 0.013%
3.3 M 0.062%