7-8. DC Voltage Amplitude Accuracy (AUX)
The DC Voltage Amplitude Accuracy test verifies the accuracy of dc voltage at the
5500A Calibrator front panel AUX terminals in the presence of a lower voltage at the
NORMAL terminals.
Nominal Value
(NORMAL)
Nominal Value
(AUX)
Measured Value
(AUX V)
Deviation %
90-Day Spec.
(% or mV)
3 V 0 mV 0.350 mV
3 V 329 mV 0.1365%
3 V -329 mV 0.1365%
3 V 0.33 V 0.1361%
3 V 3.29 V 0.0407%
3 V -3.29 V 0.0407%