5500A-SC600 Option
Verification Tables 8
8-55
8-71. Marker Generator Verification
Table 8-30. Marker Generator Verification
Period (s) Measured
Value (s)
Deviation (s) 1-Year
Spec. (s)
4.979 s 24.91E-3 s
2.002 s 4.06E-3 s
50.0 ms 3.75E-6 s
20.0 ms 50E-9 s
10.0 ms 25E-09 s
50.0 us 125E-12 s
20.0 us 50E-15 s
10.0 ns 25E-15 s
50.0 ns 125E-15 s
20.0 ns 50E-15 s
10.0 ns 25E-15 s
5.00 ns 12.5E-15 s
2.00 ns 5E-15 s
8-72. Pulse Generator Verification: Period
Table 8-31. Pulse Generator Verification: Period
Nominal
Value (V p-p)
Pulse Width
(s)
Period
(s)
Measured
Value (s)
Deviation
(s)
1-Year Spec.
(s)
2.5 8E-08 2E-06 5E-12
2.5 0.0000005 0.01 2.5E-08
2.5 0.0000005 0.02 5E-08
8-73. Pulse Generator Verification: Pulse Width
Table 8-32. Pulse Generator Verification: Pulse Width
Nominal
Value (V p-p)
Pulse Width
(s)
Period
(s)
Measured
Value (s)
Deviation
(s)
1-Year Spec.
2.5 4.0E-9 2.0E-6 2.2 ns
2.5 4.0E-9 2.0E-5 2.2 ns
2.5 4.0E-9 2.0E-4 2.2 ns
2.5 4.0E-8 2.0E-3 4 ns