Introduction and Specifications
General Specifications 1
1-11
1-16. DC Voltage Specifications
Absolute Uncertainty, tcal ± 5 °C
± (% of output + µV)
Stability
24 hours, ± 1 °C
Range
90 days 1 year
± (ppm output + µV)
Resolution
µV
Maximum
Burden
[1]
0 to 329.9999 mV 0.005 3 0.006 3 5 + 1 0.1
50
0 to 3.299999 V 0.004 5 0.005 5 4 + 3 1 10 mA
0 to 32.99999 V 0.004 50 0.005 50 4 + 30 10 10 mA
30 to 329.9999 V 0.0045 500 0.0055 500 4.5 + 300 100 5 mA
100 to 1020.000 V 0.0045 1500 0.0055 1500 4.5 + 900 1000 5 mA
Auxiliary Output (dual output mode only)
[2]
0 to 329.999 mV 0.03 350 0.04 350 30 + 100 1 5 mA
0.33 to 3.3 V 0.03 350 0.04 350 30 + 100 10 5 mA
[1] Remote sensing is not provided. Output resistance is < 5 m for outputs 0.33 V. The AUX output has an output resistance
of < 1 .
[2] Two channels of dc voltage output are provided.
Noise
Range
Bandwidth 0.1 to 10 Hz p-p
± (ppm output + µV)
Bandwidth 10 to 10 kHz rms
0 to 329.9999 mV
1 µV 4 µV
0 to 3.299999 V
10 µV 50 µV
0 to 32.99999 V
100 µV 600 µV
30 to 329.9999 V 10 ppm + 1 mV 20 mV
100 to 1020.000 V 10 ppm + 5 mV 20 mV
Auxiliary Output (dual output mode only)
[1]
0 to 329.999 mV
5 µV 20 µV
0.33 to 3.3 V
20 µV 200 µV
[1] Two channels of dc voltage output are provided.