Introduction and Specifications
Additional Specifications 1
1-29
1-34. AC Voltage, DC Offset Specifications
Range
[1]
(Normal Channel)
Offset Range
[2]
Max Peak
Signal
1-Year Absolute Offset
Uncertainty, tcal ± 5 °C
[3]
± (% Output (dc) + µV)
Sine Waves (rms)
3.3 to 32.999 mV 0 to 50 mV 80 mV 0.1 + 33
33 to 329.999 mV 0 to 500 mV 800 mV 0.1 + 330
0.33 to 3.29999 V 0 to 5 V 8 V 0.1 + 3300
3.3 to 32.9999 V 0 to 50 V 55 V 0.1 + 33 mV
Triangle Waves and Truncated Sine Waves (p-p)
9.3 to 92.999 mV 0 to 50 mV 80 mV 0.1 + 93
93 to 929.999 mV 0 to 500 mV 800 mV 0.1 + 930
0.93 to 9.29999 V 0 to 5 V 8 V 0.1 + 9300
9.3 to 92.9999 V 0 to 50 V 55 V 0.1 + 93 mV
Square Waves (p-p)
6.6 to 65.999 mV 0 to 50 mV 80 mV 0.1 + 66
66 to 659.999 mV 0 to 500 mV 800 mV 0.1 + 660
0.66 to 6.59999 V 0 to 5 V 8 V 0.1 + 6600
6.6 to 65.9999 V 0 to 50 V 55 V 0.1 + 66 mV
[1] Offsets are not allowed on ranges above the highest range shown above.
[2] The maximum offset value is determined by the difference between the peak value of the selected voltage output and the
allowable maximum peak signal. For example, a 10 V p-p square wave output has a peak value of 5 V, allowing a maximum
offset up to ± 50 V to not exceed the 55 V maximum peak signal. The maximum offset values shown above are for the minimum
outputs in each range.
[3] For frequencies 0.01 to 10 Hz, and 500 kHz to 2 MHz, the offset uncertainty is 5 % of output, ± 1 % of the offset range.
1-35. AC Voltage, Square Wave Characteristics
Rise Time
@ 1 kHz
Typical
Settling Time
@ 1 kHz
Typical
Overshoot
@ 1 kHz
Typical
Duty Cycle Range Duty Cycle Uncertainty
[1]
<1 µs
<10 µs to 1 %
of final value
<2 %
1 % to 99 %, <3.3 V p-p,
0.01 Hz to 100 kHz
± (0.8 % of period +140 ns) for
frequencies >10 kHz; + (0.8 % of
period + 2 µs) for frequencies
10 kHz.
[1] For duty cycles of 10.00 % to 90.00 %.
1-36. AC Voltage, Triangle Wave Characteristics (typical)
Linearity to 1 kHz Aberrations
0.3 % of p-p value, from 10 % to 90 % point <1 % of p-p value, with amplitude >50 % of range
1-37. AC Current (Sine Wave) Extended Bandwidth Specifications
1-Year Absolute Uncertainty,
tcal ± 5 °C,
± (% of output + % of range)
Range Frequency
% Output % Range
Maximum
Current Resolution
All current ranges, <330 mA 0.01 to 10 Hz 5.0 0.5 2 digits each range
10 Hz to 10 kHz (See AC Current (Sine Wave) Specifications)