Fluke 5500A - Leveled Sinewave Function Verification: Amplitude

Fluke 5500A
410 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A-SC300 Option
Verification Tables
8
8-91
8-118. Wave Generator Function Verification: 50
Load
Waveform
Nominal Value
(p-p)
Frequency
Measured Value
(p-p)
Deviation
(mV)
1-Year Spec.
(mV)
Square 5.0 mV 10 kHz 0.25 mV
Square 10.9 mV 10 kHz 0.43 mV
Square 44.9 mV 10 kHz 1.45 mV
Square 109.0 mV 10 kHz 3.37 mV
Square 449.0 mV 10 kHz 13.57 mV
Square 1.1 V 10 kHz 32.50 mV
Square 2.2 V 10 kHz 66.10 mV
Sine 5.0 mV 10 kHz 0.25 mV
Sine 10.9 mV 10 kHz 0.43 mV
Sine 44.9 mV 10 kHz 1.45 mV
Sine 109.0 mV 10 kHz 3.37 mV
Sine 449.0 mV 10 kHz 13.57 mV
Sine 1.1 V 10 kHz 32.50 mV
Sine 2.2 V 10 kHz 66.10 mV
Triangle 5.0 mV 10 kHz 0.25 mV
Triangle 10.9 mV 10 kHz 0.43 mV
Triangle 44.9 mV 10 kHz 1.45 mV
Triangle 109.0 mV 10 kHz 3.37 mV
Triangle 449.0 mV 10 kHz 13.57 mV
Triangle 1.1 V 10 kHz 32.50 mV
Triangle 2.2 V 10 kHz 66.10 mV
8-119. Leveled Sinewave Function Verification: Amplitude
Nominal Value (p-p) Frequency Measured Value (p-p) Deviation
(mV)
1-Year Spec.
(mV)
5.0 mV 50 kHz 0.300
10.0 mV 50 kHz 0.400
20.0 mV 50 kHz 0.600
40.0 mV 50 kHz 1.000
50.0 mV 50 kHz 1.200
100.0 mV 50 kHz 2.200
200.0 mV 50 kHz 4.200