8-48
Table 8-24. Leveled Sinewave Verification: Flatness (cont.)
Nominal
Value (V p-p)
Frequency Measured
Value (V p-p)
Deviation
(V p-p)
1-Year Spec.
(V p-p)
0.01 50 kHz na na
0.01 30 MHz 0.00025
0.01 70 MHz 0.00025
0.01 120 MHz 0.0003
0.01 290 MHz 0.0003
0.01 360 MHz 0.0005
0.01 390 MHz 0.0005
0.01 400 MHz 0.0005
0.01 480 MHz 0.0005
0.01 570 MHz 0.0005
0.01 580 MHz 0.0005
0.01 590 MHz 0.0005
0.01 600 MHz 0.0005
0.025 50 kHz na na
0.025 30 MHz 0.000475
0.025 70 MHz 0.000475
0.025 120 MHz 0.0006
0.025 290 MHz 0.0006
0.025 360 MHz 0.0011
0.025 390 MHz 0.0011
0.025 400 MHz 0.0011
0.025 480 MHz 0.0011
0.025 570 MHz 0.0011
0.025 580 MHz 0.0011
0.025 590 MHz 0.0011
0.025 600 MHz 0.0011
0.039 50 kHz na na
0.039 30 MHz 0.000685
0.039 70 MHz 0.000685
0.039 120 MHz 0.00088
0.039 290 MHz 0.00088
0.039 360 MHz 0.00166
0.039 390 MHz 0.00166
0.039 400 MHz 0.00166
0.039 480 MHz 0.00166
0.039 570 MHz 0.00166
0.039 580 MHz 0.00166
0.039 590 MHz 0.00166
0.039 600 MHz 0.00166
0.04 50 kHz na na
0.04 30 MHz 0.0007