EasyManua.ls Logo

Fluke 5500A - Page 120

Fluke 5500A
410 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Operator Manual
4-36
Note
The entered temperature will be cleared to 0
°
C (32
°
F) if you change
between tc and rtd, or change the type of thermocouple (except for a B-type
thermocouple, which clears to 600
°
C). If this should occur, select
OUTPUT tc, the desired thermocouple TYPE, and then reenter the
temperature.
OUTPUT
rtd
tc
Out; TC terminal
= 50.0900 mV
TC
MENUS
tc
TYPE
K
B
C
E
J
K
N
R
S
T
10#V/~C
TYPE
K
B
C
E
J
K
N
R
S
T
10#V/~C
UNITS
~C
REF
NONE
REF SRC
intrnl
extrnl
intrnl
~C
~F
-250 to
2316~C
Out@TC terminal (Output at the front panel TC terminals) Displays the actual dc
voltage at the front panel TC terminals. This is a display only, not a softkey function.
TC MENUS (Thermocouple Menu) Shows submenus for thermocouple outputs.

Table of Contents

Other manuals for Fluke 5500A

Related product manuals