Remote Operation
Setting up the RS-232 Host Port for Remote Control
5
5-13
4. Open Terminal from the Accessory group of Program Manager (below). If a
terminal configuration file already exists, e.g., host.trm, select the desired file
using the Open command from the File menu and go to Step 7. Otherwise, go to Step
5.
F5-0f.bmp
5. Select the Communications command from the Setting menu. Enter the RS-232
parameters that match those selected at the 5500A Calibrator for the Host port. If
using the 5500A factory defaults, the Communications dialog box for COM1 will
appear as shown below. Select COM as required. Click OK.
F5-0g.bmp
6. Verify the 5500A Calibrator is powered and in the reset condition. (If in doubt, press
the R key on the 5500A Calibrator front panel.)
7. On the Terminal screen, type the command REMOTE and press <Enter>. Observe
the 5500A Calibrator Control Display changes to REMOTE CONTROL (below).
gl41f.eps