5500A
Operator Manual
5-16
1. Turn the 5500A Calibrator power on. You may operate the calibrator during
warmup, but specifications are not guaranteed until warmup is complete.
2. Press the S key on the 5500A Calibrator front panel.
3. Negotiate the softkey selections shown below to configure the UUT serial port to
match the settings of the UUT RS-232 port. The factory defaults (shown below in
bold) are 9600 baud, 8 data bits, 1 stop bit, and no parity. Other parameters include
flow control (STALL).
gl45f.eps