5500A
Operator Manual
5-20
6. At the prompt, type the following line to activate the IEEE interface card:
<ibdev 0 4 0 10 1 0>
The second number in this line is the primary address of the calibrator. If the address
has been changed from the factory default, change this line accordingly.
7. The prompt reads
<ud0:>. From this prompt, type
<ibwrt "uut_sendb 82,69,77,83,11,13">
8. Press the ENTER (or RETURN) key. This command will send REMS<CR><LF> to
the UUT serial port. After the command is entered, the Win32 Interactive Control
shows the status of the command. If an error is encountered, check the typing or
consult the National Instruments manual regarding Win32 Interactive control. The
count message is the amount of characters sent over the bus.
9. Verify that the UUT is in remote.
10. From the ud0: prompt type <q> then press the ENTER (or RETURN) key.