Remote Commands
Command Summary by Function
6
6-5
Output Commands
BOOST Activates and deactivates a 5725A Amplifier and sets the source preference to the
5725A Amplifier or to the 5500A Calibrator.
BOOST? Returns whether or not the 5725A Amplifier is in use and which output is being
boosted.
CFREQ? Returns the optimum frequency value for stimulus for capacitance modes.
DC_OFFSET Applies a DC offset to an AC output voltage.
DC_OFFSET? Returns the DC offset voltage.
DPF Sets the displacement power factor (phase angle) between the NORMAL and AUX
terminals for AC power output only.
DPF? Returns the displacement power factor (phase angle) between the NORMAL and
AUX terminals.
DUTY Sets the duty cycle of SQUARE wave outputs.
DUTY? Returns the duty cycle.
FUNC? Returns the present output, measurement, or calibration function.
HARMONIC Make the frequency of one output be a harmonic (multiple) of the other output (called
the fundamental).
HARMONIC? Returns the present instrument harmonic and fundamental locations.
OPER Activates the 5500A output if it is in standby.
OPER? Returns the operate/standby setting.
OUT Sets the output of the 5500A and establishes a new reference point for the error
mode.
OUT? Returns the output amplitudes and frequency of the 5500A.
PHASE Sets the phase between the NORMAL and AUX terminals for dual outputs. The
NORMAL terminal output is the phase reference.
PHASE? Returns the phase between the NORMAL and AUX terminals.
POWER? Returns the equivalent power for DC and AC power output.
RANGE? Returns the present output ranges.
RANGELCK Locks in the present range, or selects auto ranging.
RANGELCK? Returns whether or not the preset output range is locked.
STBY Puts the 5500A in standby.
WAVE Sets the waveforms for AC outputs.
WAVE? Returns the waveforms of the output.
ZCOMP Activate (2-wire or 4-wire) or deactivate impedance compensation.
ZCOMP? Returns whether or not impedance compensation is active and if active, which type.