EasyManua.ls Logo

Fluke 5500A - Adjusting the Leveled Sine Wave Function

Fluke 5500A
410 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Operator Manual
6-42
TC_TYPE_D?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Thermocouple Type Default query) Return the default thermocouple (TC) sensor type.
Responses: B (B-type thermocouple)
C (C-type thermocouple)
E (E-type thermocouple)
J (J-type thermocouple)
K (K-type thermocouple) (default)
N (N-type thermocouple)
R (R-type thermocouple)
S (S-type thermocouple)
T (T-type thermocouple)
X (10 µV/°C linear output)
Example: TC_TYPE_D? returns K
Return K when the thermocouple type default is a type-K thermocouple.
TEMP_STD
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Temperature Degree Standard command) Select the temperature standard ipts-68 (1968
International Provisional Temperature Standard) or its-90 (1990 International
Temperature Standard), which is saved in the 5500A non-volatile memory. (While
saving configuration data in the non-volatile memory, a period of about 2 seconds, the
5500A does not respond to remote commands.) The default is its-90.
Parameters: IPTS_68
ITS_90
Example: TEMP_STD ITS-90
See the temperature standard to its-90.
TEMP_STD?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Temperature Degree Standard command) Return the temperature standard ipts-68
(1968 International Provisional Temperature Standard) or its-90 (1990 International
Temperature Standard).
Responses: IPTS_68
ITS_90
Example: TEMP_STD? returns ITS-90
Return ITS-90 when the temperature degree standard is the 1990 International
Temperature Standard.