5500A
Operator Manual
6-44
*TST?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Self Test command) Initiate self-test and return a 0 for pass or a 1 for fail. If any faults
are detected, they are displayed on screen (terminal mode) or are logged into the fault
queue where they can be read by the ERR? query (computer mode).
Response: 0 (pass self test)
1 (fail self test)
Example: *TST? returns 1
Return 1 when self test is successful.
UUT_FLUSH
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Flush UUT Receive Buffer command) Flush the UUT receive buffer for data received
from the UUT over the 5500A Calibrator rear panel SERIAL 2 TO UUT serial port. The
command may be sent over gpib or RS-232 ports, but applies to SERIAL 2 TO UUT
serial port operation.
Parameter: (None)
Example: UUT_FLUSH
Flush the 5500A Calibrator receive data buffer for the UUT.
UUT_RECV?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Return UUT Receive Data command) Return data from the UUT in IEEE-488.2
Standard format over the 5500A Calibrator rear panel SERIAL 2 TO UUT serial port.
The command may be sent over gpib or RS-232 ports, but applies to SERIAL 2 TO UUT
serial port operation.
Response: <data> (binary block data in definite length format from UUT)
Example: UUT_RECV? returns #211+1.99975E+0
Return (for example) a measurement from the UUT. The format is #2 (two numbers
follow) 11 (characters follow) +1.99975E+0 (11 characters).