5500A
Operator Manual
8-52
Table 8-24. Leveled Sinewave Verification: Flatness (cont.)
Nominal
Value (V p-p)
Frequency Measured
Value (V p-p)
Deviation
(V p-p)
1-Year Spec.
(V p-p)
1.3 480 MHz 0.0521
1.3 570 MHz 0.0521
1.3 580 MHz 0.0521
1.3 590 MHz 0.0521
1.3 600 MHz 0.0521
3.4 50 kHz na na
3.4 30 MHz 0.0511
3.4 70 MHz 0.0511
3.4 120 MHz 0.0681
3.4 290 MHz 0.0681
3.4 360 MHz 0.1361
3.4 390 MHz 0.1361
3.4 400 MHz 0.1361
3.4 480 MHz 0.1361
3.4 570 MHz 0.1361
3.4 580 MHz 0.1361
3.4 590 MHz 0.1361
3.4 600 MHz 0.1361
5.5 50 kHz na na
5.5 30 MHz 0.0826
5.5 70 MHz 0.0826
5.5 120 MHz 0.1101
5.5 290 MHz 0.1101
5.5 360 MHz 0.2201
5.5 390 MHz 0.2201
5.5 400 MHz 0.2201
5.5 480 MHz 0.2201
5.5 570 MHz 0.2201
5.5 580 MHz 0.2201
5.5 590 MHz 0.2201
5.5 600 MHz 0.2201