5500A
Service Manual
6-54
Table 6-41. Wave Generator Verification at 1 M
Calibrator
Mainframe
Wave Type
Calibrator
Mainframe
output
(@ 10 kHz)
5790A
Reading
(V rms)
Conversion
Factor
5790A Reading x
Conversion Factor
(V p-p)
Tolerance
(V p-p)
square 1.8 mV 2.0000 0.000154 V
square 11.9 mV 2.0000 0.000457 V
square 21.9 mV 2.0000 0.00075 V
square 22.0 mV 2.0000 0.00076 V
square 56.0 mV 2.0000 0.00178 V
square 89.9 mV 2.0000 0.002797 V
square 90 mV 2.0000 0.0028 V
square 155 mV 2.0000 0.00475 V
square 219 mV 2.0000 0.00667 V
square 220 mV 2.0000 0.0067 V
square 560 mV 2.0000 0.0169 V
square 899 mV 2.0000 0.02707 V
square 0.90 V 2.0000 0.0271 V
square 3.75 V 2.0000 0.1126 V
square 6.59 V 2.0000 0.1978 V
square 6.6 V 2.0000 0.1981 V
square 30.8 V 2.0000 0.9241 V
square 55.0 V 2.0000 1.6501 V
sine 1.8 mV 2.8284 0.000154 V
sine 21.9 mV 2.8284 0.000757 V
sine 89.9 mV 2.8284 0.002797 V
sine 219 mV 2.8284 0.00667 V
sine 899 mV 2.8284 0.02707 V
sine 6.59 V 2.8284 0.1978 V
sine 55 V 2.8284 1.6501 V
triangle 1.8 mV 3.4641 0.000154 V
triangle 21.9 mV 3.4641 0.000757 V
triangle 89.9 mV 3.4641 0.002797 V
triangle 219 mV 3.4641 0.00667 V
triangle 899 mV 3.4641 0.02707 V
triangle 6.59 V 3.4641 0.1978 V
triangle 55 V 3.4641 1.6501 V