SC600 Option
Verification
6
6-55
Table 6-42. Wave Generator Verification at 50
Calibrator
Mainframe
Wave
Type
Calibrator
Mainframe
output
(10 kHz)
5790A
Reading
(V rms)
Conversion
Factor
5790A Rdg x
Conversion
Factor (V p-p)
V p-p value x
correction
Tolerance
(V p-p)
square 1.8 mV 2.0000 0.000154 V
square 6.4 mV 2.0000 0.000292 V
square 10.9 mV 2.0000 0.000427 V
square 11.0 mV 2.0000 0.00043 V
square 28.0 mV 2.0000 0.00094 V
square 44.9 mV 2.0000 0.001447 V
square 45 mV 2.0000 0.00145 V
square 78 mV 2.0000 0.00244 V
square 109 mV 2.0000 0.00337 V
square 110 mV 2.0000 0.0034 V
square 280 mV 2.0000 0.0085 V
square 449 mV 2.0000 0.01357 V
square 450 mV 2.0000 0.0136 V
square 780 mV 2.0000 0.0235 V
square 1.09 V 2.0000 0.0328 V
square 1.10 V 2.0000 0.0331 V
square 1.80 V 2.0000 0.0541 V
square 2.50 V 2.0000 0.0751 V
sine 1.8 mV 2.8284 0.000154 V
sine 10.9 mV 2.8284 0.000427 V
sine 44.9 mV 2.8284 0.001447 V
sine 109 mV 2.8284 0.00337 V
sine 449 mV 2.8284 0.01357 V
sine 1.09 V 2.8284 0.0328 V
sine 2.50 V 2.8284 0.0751 V
triangle 1.8 mV 3.4641 0.000154 V
triangle 10.9 mV 3.4641 0.000427 V
triangle 44.9 mV 3.4641 0.001447 V
triangle 109 mV 3.4641 0.00337 V
triangle 449 mV 3.4641 0.01357 V
triangle 1.09 V 3.4641 0.0328 V
triangle 2.50 V 3.4641 0.0751 V