EasyManuals Logo

Fluke 5700A Series II User Manual

Fluke 5700A Series II
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #103 background imageLoading...
Page #103 background image
Front Panel Operation
Connecting the Calibrator to a UUT 4
4-17
UUT
HI
LO
A
HI
LO
GUARD
INPUT
SENSE
4-WIRE
HI
HI
LO
LO
HI
OUTPUT
V A
SENSE
V
AUX
CURRENT
GUARD GROUND
WIDEBAND
CALIBRATOR
UUT
HI
HI
LO
LO
HI
OUTPUT
V A
SENSE
V
AUX
CURRENT
GUARD GROUND
WIDEBAND
UUT
HI
HI
LO
LO
HI
OUTPUT
V A
SENSE
V
AUX
CURRENT
GUARD GROUND
WIDEBAND
C.
B.
A.
EX SNS
: ON
EX SNS
: OFF
EX GRD
: OFF
EX SNS
: OFF
EX GRD
: ON
DC
INPUT
AC
LOHI
EX SNS
: ON
SHIELDED TWISTED PAIR
SHIELDED TWISTED PAIR
CALIBRATOR
CALIBRATOR
Figure 4-1. UUT Connections: DC Voltage, AC Voltage 10 kHz

Table of Contents

Other manuals for Fluke 5700A Series II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series II and is the answer not in the manual?

Fluke 5700A Series II Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0 °C to 50 °C
Storage Temperature-20 °C to 70 °C
Voltage Range0 to 1100 V
DC Voltage Range10 μV to 1100 V
Frequency Range1 MHz
Humiditynon-condensing
Power Requirements100 V to 240 V

Related product manuals