EasyManuals Logo

Fluke 5700A Series II User Manual

Fluke 5700A Series II
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #163 background imageLoading...
Page #163 background image
Remote Operation
Commands 5
5-13
Table 5-1. Command Summary by Function
Error Mode Commands
ADJOUT? Returns the adjusted output magnitude and frequency
INCR Increments or decrements the output
MULT Multiplies the reference by a value and establishes a new reference
NEWREF Establishes a new reference
OFFSET Sets and enables or disables an offset
OFFSET? Returns the value of the present offset
OLDREF Sets the output to the present reference
OUT_ERR? Returns the UUT error computed by the calibrator
REFOUT? Returns the value of the present reference
SCALE Activates and deactivates scaling
SCALE? Returns scaling information
SCAL_ERR? Returns the value of the scale error if scaling is active
Instrument Configuration Commands
BTYPE Selects the amplifier type for voltage or current boost
BTYPE? Returns the amplifier types set for voltage and current boost
CLOCK Sets the clock/calendar
CLOCK? Returns the setting of the clock/calendar
CUR_POST Selects the active binding post for current output
CUR_POST? Returns the active binding post for current output
EXTGUARD Opens and closes an internal connection between GUARD and LO
EXTSENSE Opens and closes an internal connection between SENSE and OUTPUT
FORMAT Restores calibration constants and setups to factory defaults
LIMIT Sets positive and negative output limits
LIMIT? Returns the programmed positive and negative output limits
*LRN? Returns a list of commands that would duplicate the present state
RANGE? Returns the present output range
RANGELCK Locks in the present range, or selects autoranging

Table of Contents

Other manuals for Fluke 5700A Series II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series II and is the answer not in the manual?

Fluke 5700A Series II Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0 °C to 50 °C
Storage Temperature-20 °C to 70 °C
Voltage Range0 to 1100 V
DC Voltage Range10 μV to 1100 V
Frequency Range1 MHz
Humiditynon-condensing
Power Requirements100 V to 240 V

Related product manuals