EasyManua.ls Logo

Fluke 5700A Series II - Page 180

Fluke 5700A Series II
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5700A/5720A Series II
Operators Manual
5-30
Table 5-2. Commands (continued)
*IDN?
Description Sequential command.
Returns instrument model number and firmware version letters for the main, inguard, and if
attached, the 5725A Amplifier CPU.
Parameter None
Response (String, string, 0, string)
A message containing four fields separated by commas, as follows:
1. Fluke
2. Model number (5700A or 5720A)
3. Serial number
4. Three firmware versions: one each for the Main CPU, the Inguard CPU, and the Boost
CPU. Each version is separated by a plus (+). If no amplifier is attached then its position
(third character) contains an asterisk (*).
Example: FLUKE,5720A,5248000,1.0+B+*
INCR
Description Overlapped command, ignored if not in remote.
Increments the output amplitude and enters error mode, the same as using the output
adjustment knob in local operation.
Parameter Increment step size; positive for incremental step, negative for decremental step. Units
(optional) specify magnitude or frequency.
Examples INCR -.00001 Enters error mode and decrements the output by .00001.
INCR 1 Hz Enters error mode and increments the frequency by 1 Hz.
ISCE
Description Sequential command.
Loads a byte into the Instrument Status Change Enable register described under "Checking
the Calibrator’s Status."
Parameter The decimal equivalent of the binary number to load into the register.
Example: ISCE 56
Enables bits 3 (BOOST), 4 (RCOMP), and 5 (RLOCK) in the Service Request Enable
register.
ISCE?
Description Sequential command.
Returns the byte from the Instrument Status Change Enable register, described under
"Checking the Calibrator’s Status."
Parameter None
Response The decimal equivalent of the register contents byte.
Example: ISCE?
Returns: "4" if bit 3 (BOOST) is enabled (1) and the rest of the bits are disabled (0). (See
"Checking the Calibrator’s Status" for details.)