Remote Operation
Commands 5
5-31
Table 5-2. Commands (continued)
ISCR?
Description Sequential command.
Returns and clears the byte from the Instrument Status Change Register, described under
"Checking the Calibrator’s Status."
Parameter None
Response The decimal equivalent of the register contents byte.
Example: ISCR?
Returns: "8" if bit 3 (BOOST) is set (1) and the rest of the bits are reset (0). (See "Checking
the Calibrator’s Status" for details.)
ISR?
Description Sequential command.
Returns and clears the byte from the Instrument Status Register, described under
"Checking the Calibrator’s Status."
Parameter None
Response The decimal equivalent of the register contents byte.
Example: ISR?
Returns: "16" if bit 4 (RLOCK) is set (1) and the rest of the bits are reset (0). (See "Checking
the Calibrator’s Status" for details.)
LIMIT
Description Sequential command, ignored if not in remote.
Sets the maximum permissible output magnitudes, negative and positive. Both parameters
must be present, and the units must be none, volts, or amps. The units must not conflict. If
no units are specified, V is assumed.
Parameter
1. Positive limit value with optional multiplier and optional unit.
2. Negative limit value with optional multiplier and optional unit.
Example: LIMIT 1.8A, -1.2A
Sets the limit for ac and positive dc current to 1.8A, and negative dc current to -1.2A.
LIMIT?
Description Sequential command.
Returns the programmed output magnitude limits for voltage and current.
Response (Float, float, float, float)
1. Positive voltage limit value
2. Negative voltage limit value
3. Positive current limit value
4. Negative current limit value
Example: LIMIT?
Returns: "2.20E-2,-1.0E-2,1.8E0,-1.2E0" if the programmed voltage range is -100 to 220V,
and the programmed current range is -1.2 to 1.8A.