5700A/5720A Series II
Operators Manual
5-32
Table 5-2. Commands (continued)
*LRN?
Description Sequential command.
Provides the current setting of the calibrator. The response to this command is a string that
recreates the state of the following settings when returned to the calibrator:
Output value Whether the wideband AC Module is on Phase lock
Operate/standby Whether an auxiliary amplifier is on Phase shift
Guard Range lock Offset
Sense Two-wire compensation Scale
Output limits
Parameter None
Response (String) A string that enables the settings that were set at the time of the *LRN? command.
Example:
PRINT @4, "*LRN?"
INPUTLINE @4, A$
.
.
(Later, to recreate the settings:)
PRINT @4, A$
MULT
Description Overlapped command, ignored if not in remote.
Multiplies the reference magnitude by the parameter and changes the output to the new
value. The reference magnitude is the present output in direct operation, and the reference
in error mode.
Parameter Floating-point number to act as a multiplier
Example: If the output is 10V with error mode off and you send the command, "MULT 1.9", the output
changes to 19V.
NEWREF
Description Sequential command, ignored if not in remote.
Sets reference value to be the present output value. (The same as pressing the NEW REF
key in local operation.)
Parameter None
OFFSET
Description Sequential command, ignored if not in remote.
Enables and disables an offset for the output value. Effective immediately when enabled.
Parameters: ON or OFF
OFFSET?
Description Sequential command.
Returns the offset value if the offset is enabled; otherwise, returns the number 0.0.
Parameter None
Response 1. (Float) Offset value or 0.0 if no offset is active
2. (Float) Offset units
Example: OFFSET?
Returns: "5.05000E-03,V" if an offset of 5.05 mV is active.