5700A/5720A Series II
Operators Manual
5-36
Table 5-2. Commands (continued)
*PUD?
Description Sequential command.
Returns the contents of the *PUD (Protected User Data) memory.
Parameter None
Response #(non-zero digit) (digits) (user data)
The non-zero digit specifies the number of characters that will follow in the \<digits\> field.
These characters are 0 through 9 (ASCII 48 through 57 decimal). The value of the number
in the \<digits\> field defines the number of user data bytes that follow in the \<user data\>
field. The maximum response is 64 characters.
Example *PUD?
Returns: "205FLUKE" assuming that this is stored as in the example for PUD* above.
RANGE?
Description Sequential command.
Returns the present output range.
Parameter None
Response (String) Symbolic name of the range from Table 5-4.
RANGELCK
Description Overlapped command, ignored if not in remote.
Locks or unlocks the present output range. The range automatically unlocks if the output
function changes, for example from dc volts to dc current.
Parameter
ON (Locks the range)
OFF (Unlocks the range)
RCOMP
Description Overlapped command, ignored if not in remote.
While a resistance output 19 k or lower is selected, RCOMP activates or deactivates two-
wire ohms compensation circuitry.
Parameter
ON (Turns on the two-wire compensation circuitry)
OFF (Turns off the two-wire compensation circuitry)
REFOUT?
Description Sequential command.
Returns the value of the reference, which is the output value of the calibrator the last time a
new reference was established with an OUT, NEW_REF, or MULT.
Parameter None
Response
1. (Float) Output amplitude
2. (String) Units (V, A, OHM, or DBM)
3. (String) Frequency (0 if dc)