Remote Operation
Commands 5
5-37
Table 5-2. Commands (continued)
RPT_STR
Description Sequential command.
Loads the user report string. The user report string can be read on the Control Display in
local operation, and appears on calibration reports. The CALIBRATION switch must be set
to ENABLE.
Parameter String of up to 40 characters
RPT_STR?
Description Sequential command.
Returns the user report string. The user report string can be read on the Control Display in
local operation, and appears on calibration reports.
Parameter None
Response (String) Up to 40 characters
*RST
Description Overlapped command, ignored if not in remote.
(Reset.) Sets the calibrator to its power-up default state: 0V, 0 Hz, Standby, [BOOST] off,
[W BND] off, [EX GRD] off, [EX SNS] off, phase lock off, phase shift off, range lock off,
[SCALE] off, [OFFSET] off, two-wire compensation off, and entry limits set to defaults.
*RST does not affect any of the following:
State of the IEEE-488 interface
Selected bus address
Status Register Enable setting
Contents of nonvolatile memory
SCALE
Description Sequential command, ignored if not in remote.
Turns scaling on or off. (The same as pressing the SCALE key on the front panel.)
Parameters: ON or OFF
SCALE?
Description Sequential command.
Returns the full-scale nominal and full-scale actual values set when SCALE more was
turned on. If scaling is off, both responses one and two are "0.0."
Responses:
1. (Float) Nominal full-scale value
2. (Float) Actual full-scale value
3. (String) V, A, or DBM; units for the first two responses