EasyManuals Logo

Fluke 5700A Series II User Manual

Fluke 5700A Series II
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #236 background imageLoading...
Page #236 background image
5700A/5720A Series II
Operators Manual
6-12
119.99 kHz 0.0018 V 0.0008 % 0.20 % 0.40
0.1200 MHz 0.0019 V 0.0009 % 0.20 % 0.44
0.2000 MHz 0.0040 V 0.0018 % 0.20 % 0.92
0.3000 MHz 0.0142 V 0.0064 % 0.20 % 3.22
0.4000 MHz 0.0288 V 0.0131 % 0.57 % 2.29
0.5000 MHz 0.0266 V 0.0121 % 0.57 % 2.12
0.6000 MHz 0.0506 V 0.0230 % 1.35 % 1.70
0.7000 MHz 0.0613 V 0.0279 % 1.35 % 2.07
0.8000 MHz 0.0546 V 0.0248 % 1.35 % 1.84
0.9000 MHz 0.0580 V 0.0264 % 1.35 % 1.95
1.0000 MHz 0.0782 V 0.0355 % 1.35 % 2.63
1100V FS 0.000 V 0.14 ppm 83.64 ppm 0.17
0.3000 kHz -0.000 V -0.29 ppm 83.64 ppm -0.35
0.5000 kHz -0.009 V -8.54 ppm 83.64 ppm -10.21
0.7000 kHz -0.004 V -3.45 ppm 83.64 ppm -4.13
1.0000 kHz -0.003 V -2.41 ppm 83.64 ppm -2.88
-----------------------------------------------------------------------------
DC CURRENT OUTPUT SHIFTS
-----------------------------------------------------------------------------
RANGE Point Zero Shift Full Scale Shift Spec Shift
(+/-) (% spec)
220 uA +FS -0.0005 uA -0.0003 uA -1.17 ppm 95.45 ppm -1.23
-FS -0.0005 uA -0.0007 uA -3.15 ppm 95.45 ppm -3.30
2.2 mA +FS -0.000005 mA -0.000004 mA -1.70 ppm 54.55 ppm -3.11
-FS -0.000005 mA -0.000005 mA -2.47 ppm 54.55 ppm -4.53
22 mA +FS -0.00005 mA -0.00001 mA -0.63 ppm 54.55 ppm -1.15
-FS -0.00004 mA -0.00007 mA -3.33 ppm 54.55 ppm -6.11
220 mA +FS -0.0005 mA -0.0005 mA -2.48 ppm 64.55 ppm -3.84
-FS -0.0005 mA -0.0004 mA -1.82 ppm 64.55 ppm -2.81
2.2A +FS -0.000005 A -0.000022 A -10.01 ppm 93.64 ppm -10.70
-FS -0.000005 A 0.000012 A 5.57 ppm 93.64 ppm 5.95
-----------------------------------------------------------------------------
AC CURRENT OUTPUT SHIFTS
-----------------------------------------------------------------------------
RANGE Point Full Scale Shift Spec Shift
(+/-) (% spec)
220 uA FS 0.000 uA 0.97 ppm 230.91 ppm 0.42
0.5000 kHz 0.002 uA 8.31 ppm 230.91 ppm 3.60
1.0000 kHz -0.000 uA -0.40 ppm 230.91 ppm -0.17
2.000 kHz -0.001 uA -0.0003 % 0.53 % -0.05
5.000 kHz 0.000 uA 0.0001 % 0.53 % 0.02
7.000 kHz -0.001 uA -0.0003 % 2.26 % -0.01
10.000 kHz -0.002 uA -0.0008 % 2.26 % -0.04
2.2 mA FS 0.00000 mA 0.37 ppm 158.18 ppm 0.23
0.5000 kHz 0.00001 mA 2.67 ppm 158.18 ppm 1.69
1.0000 kHz -0.00001 mA -2.48 ppm 158.18 ppm -1.57
2.000 kHz 0.00000 mA 0.0002 % 0.17 % 0.11
5.000 kHz -0.00001 mA -0.0006 % 0.17 % -0.34
7.000 kHz -0.00001 mA -0.0002 % 0.62 % -0.04
10.000 kHz -0.00001 mA -0.0004 % 0.62 % -0.07
22 mA FS 0.0000 mA 1.33 ppm 158.18 ppm 0.84
0.5000 kHz 0.0000 mA 0.40 ppm 158.18 ppm 0.25
Figure 6-2. Calibration Shift Results Printout (continued)

Table of Contents

Other manuals for Fluke 5700A Series II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series II and is the answer not in the manual?

Fluke 5700A Series II Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0 °C to 50 °C
Storage Temperature-20 °C to 70 °C
Voltage Range0 to 1100 V
DC Voltage Range10 μV to 1100 V
Frequency Range1 MHz
Humiditynon-condensing
Power Requirements100 V to 240 V

Related product manuals