5700A/5720A Series II
Operators Manual
6-12
119.99 kHz 0.0018 V 0.0008 % 0.20 % 0.40
0.1200 MHz 0.0019 V 0.0009 % 0.20 % 0.44
0.2000 MHz 0.0040 V 0.0018 % 0.20 % 0.92
0.3000 MHz 0.0142 V 0.0064 % 0.20 % 3.22
0.4000 MHz 0.0288 V 0.0131 % 0.57 % 2.29
0.5000 MHz 0.0266 V 0.0121 % 0.57 % 2.12
0.6000 MHz 0.0506 V 0.0230 % 1.35 % 1.70
0.7000 MHz 0.0613 V 0.0279 % 1.35 % 2.07
0.8000 MHz 0.0546 V 0.0248 % 1.35 % 1.84
0.9000 MHz 0.0580 V 0.0264 % 1.35 % 1.95
1.0000 MHz 0.0782 V 0.0355 % 1.35 % 2.63
1100V FS 0.000 V 0.14 ppm 83.64 ppm 0.17
0.3000 kHz -0.000 V -0.29 ppm 83.64 ppm -0.35
0.5000 kHz -0.009 V -8.54 ppm 83.64 ppm -10.21
0.7000 kHz -0.004 V -3.45 ppm 83.64 ppm -4.13
1.0000 kHz -0.003 V -2.41 ppm 83.64 ppm -2.88
-----------------------------------------------------------------------------
DC CURRENT OUTPUT SHIFTS
-----------------------------------------------------------------------------
RANGE Point Zero Shift Full Scale Shift Spec Shift
(+/-) (% spec)
220 uA +FS -0.0005 uA -0.0003 uA -1.17 ppm 95.45 ppm -1.23
-FS -0.0005 uA -0.0007 uA -3.15 ppm 95.45 ppm -3.30
2.2 mA +FS -0.000005 mA -0.000004 mA -1.70 ppm 54.55 ppm -3.11
-FS -0.000005 mA -0.000005 mA -2.47 ppm 54.55 ppm -4.53
22 mA +FS -0.00005 mA -0.00001 mA -0.63 ppm 54.55 ppm -1.15
-FS -0.00004 mA -0.00007 mA -3.33 ppm 54.55 ppm -6.11
220 mA +FS -0.0005 mA -0.0005 mA -2.48 ppm 64.55 ppm -3.84
-FS -0.0005 mA -0.0004 mA -1.82 ppm 64.55 ppm -2.81
2.2A +FS -0.000005 A -0.000022 A -10.01 ppm 93.64 ppm -10.70
-FS -0.000005 A 0.000012 A 5.57 ppm 93.64 ppm 5.95
-----------------------------------------------------------------------------
AC CURRENT OUTPUT SHIFTS
-----------------------------------------------------------------------------
RANGE Point Full Scale Shift Spec Shift
(+/-) (% spec)
220 uA FS 0.000 uA 0.97 ppm 230.91 ppm 0.42
0.5000 kHz 0.002 uA 8.31 ppm 230.91 ppm 3.60
1.0000 kHz -0.000 uA -0.40 ppm 230.91 ppm -0.17
2.000 kHz -0.001 uA -0.0003 % 0.53 % -0.05
5.000 kHz 0.000 uA 0.0001 % 0.53 % 0.02
7.000 kHz -0.001 uA -0.0003 % 2.26 % -0.01
10.000 kHz -0.002 uA -0.0008 % 2.26 % -0.04
2.2 mA FS 0.00000 mA 0.37 ppm 158.18 ppm 0.23
0.5000 kHz 0.00001 mA 2.67 ppm 158.18 ppm 1.69
1.0000 kHz -0.00001 mA -2.48 ppm 158.18 ppm -1.57
2.000 kHz 0.00000 mA 0.0002 % 0.17 % 0.11
5.000 kHz -0.00001 mA -0.0006 % 0.17 % -0.34
7.000 kHz -0.00001 mA -0.0002 % 0.62 % -0.04
10.000 kHz -0.00001 mA -0.0004 % 0.62 % -0.07
22 mA FS 0.0000 mA 1.33 ppm 158.18 ppm 0.84
0.5000 kHz 0.0000 mA 0.40 ppm 158.18 ppm 0.25
Figure 6-2. Calibration Shift Results Printout (continued)