EasyManua.ls Logo

Fluke 5700A Series II - Page 80

Fluke 5700A Series II
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5700A/5720A Series II
Operators Manual
3-10
Table 3-1. Front Panel Features (continued)
a
WIDEBAND Connector
A Type "N" connector that provides a connection point for output from the
Option 5700A-03 Wideband AC Module. Wideband output specifications are
stated for output levels present at the end of its 3-foot 50 coaxial cable
terminated into a 50 purely resistive load. The connector shell is connected
to chassis ground. Refer to Chapter 4 for connecting and operating
instructions for the wideband module.
b
GND Binding Post
If the calibrator is the location of the ground reference point in a system, the
GND binding post can be used for connecting other instruments to earth
ground. (The chassis is normally connected to earth ground through the
three-conductor line cord instead of through the earth ground binding post.)
Refer to "Cable Connections" in Chapter 4 for details. Included with the
calibrator is a brass strap that connects GND to V GUARD.
c
V GUARD Binding Post
Provides an external connection point for the internal voltage guard. For a
UUT with floating (ungrounded) inputs, the V GUARD should be connected
to LO internally (
x off). For a UUT with a grounded input, the V GUARD
must be externally connected to the grounded UUT input (x on). The
maximum allowable potential between the V GUARD connector and chassis
ground is 20V peak. Refer to "When to use External Voltage Guard" and
"Cable Connections" in Chapter 4 for details.
d
SENSE Binding Posts
Used in the resistance and voltage functions for sensing at the UUT after
you have selected external sense by pressing
X or by remote command.
External Sensing should be used in the dc voltage function when the UUT
draws enough current to produce a significant voltage drop in the cables,
and in the resistance function when the UUT has a four-wire ohms input and
the calibrator is set to 100 k or less. External sensing can also be used in
the two-wire ohms function to allow the two-wire compensation circuitry to
the UUT terminals. Refer to Refer to "When to use External Sensing," "Four-
Wire vs. Two-Wire Resistance Connections," and "Cable Connections" in
Chapter 4 for external sensing instructions and illustrations of SENSE
connections.
e
OUTPUT Binding Posts
Provide connection points for ac and dc current and voltage output, as well
as resistance. The function of each OUTPUT binding post is defined below:
LO
The common binding post for all output functions including 5725A amplified
voltage output, but not Option 5700A-03 Wideband AC or other auxiliary
amplifier output.
HI
The active binding post for all output functions including 5725A amplified
voltage output, but not Option 5700A-03 Wideband AC or other auxiliary
amplifier output.
AUX CURRENT OUTPUT
An optional active binding post for current. It is convenient to use the AUX
CURRENT OUTPUT binding post when calibrating a UUT with a separate
current input terminal. Refer to "Connecting the calibrator to the UUT" in
Chapter 4 for instructions for using this binding post.

Table of Contents

Other manuals for Fluke 5700A Series II

Related product manuals