Fluke DSX-600 - Set the NVP to a Specified Value; Find a Cables Actual NVP

Fluke DSX-600
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DSX-600 CableAnalyzer
Technical Reference Handbook
186
6 Tap MEASURE.
7 On the NVP screen, use or to change the NVP value
until the measured length is the length of the cable, then tap
SAVE.
Reset the NVP to the Default Value
To reset the NVP to the default value defined by the selected
cable type, select the cable type again on the TEST SETUP screen.
Make a Custom Outlet Configuration
Custom outlet configurations start with the T568A or T568B
configurations. You can enable or disable each pair, and give the
configuration a name.
To make a custom outlet configuration
1 On the home screen, tap the test setup panel, tap a copper
test, then tap EDIT.
2 On the TEST SETUP screen, tap Outlet Configuration.
3 Tap CUSTOM, then tap MANAGE, then tap the Create panel.
4 Select the T568A or T568B configuration then select the pairs
you want enabled.
5 Enter a name for the configuration, then tap SAVE. The tester
adds an asterisk (*) to the start and end of the custom
configuration name.
6 To use the configuration, tap it on the OUTLET CONFIG
screen, then tap USE SELECTED.