Chapter 12: Specifications
Performance Specifications
207
Dynamic
Accuracy
ACR-F
± 1.0 dB (FEXT dynamic accuracy is tested to ± 0.75 dB)
Source/load
return loss
1to 300MHz:
20 – 12.5log(f/100),
20 dB maximum
300 to 500 MHz:
14 dB
1 to 300 MHz: 18 – 12.5 log(f/100),
20 dB maximum
300to 500MHz: 12dB
Random
Noise Floor
95 dB 90 dB
Residual
NEXT
90 – 20 log(f/100)
(measured to 85 dB
maximum)
85 - 20 log(f/100)
(measured to 85 dB
maximum)
1 to 500 MHz:
82.3 - 20 log(f/100)
(measured to 85 dB
maximum)
Residual FEXT 80 - 20 log(f/100)
(measured to 85 dB
maximum)
75 - 20 log(f/100)
(measured to 85 dB
maximum)
60 - 15 log(f/100)
(measured to 85 dB
maximum)
Output
Signal
Balance
1 to 500 MHz:
40 - 20 log(f/100)
(measured to 60 dB
maximum)
1 to 500 MHz: 37 - 20 log(f/100)
(measured to 60 dB maximum)
Common
Mode
Rejection
40 - 20 log(f/100)
(measured to 60 dB
maximum)
37 - 20 log(f/100)
(measured to 60 dB maximum)
Tracking 0.5 dB (applicable when IL > 3 dB)