Insert collection tubes
Insert a sufficient number of collection tubes in to the Tube rack, starting at position 1, pushing each
one down as far as they will go. All the tubes must be of the same length and diameter and there
should be no spaces in the sequence.
Adjust the Delivery arm
Follow the instructions in the table below to adjust the height of the Delivery arm.
ActionStep
Lift and then lower the Delivery arm (1), and allow it to move in so the Tube sensor (2)
touches the collection tubes of the outer track.
1
1
2
Loosen the lock knob (3)2
3
204 ÄKTA pure User Manual 29119969 AB
5 Operation
5.8 Fractionation
5.8.2 Prepare Fraction collector F9-R