Possible cause and actionProblem
The waste tubing is positioned so that the flow is obstructed
Untwist the waste tubing.
The waste container is placed in a position higher than the
waste outlet
Place the waste container in a position lower than the waste
outlet.
The waste tubing is blocked.
Clean or replace the waste tubing.
The waste or the interior of
the fraction collector is
flooded
Check the position of the waste funnel
Refit the waste funnel.
The liquid leaving the nozzle
does not strike the waste
funnel
Salt residuals in the tubing
Perform a Fraction collector wash, see Section 7.6.4 Clean
Fraction collector F9-C, on page 288.
If persistent, contact Service.
The tubing in the fraction
collector is blocked
Leakage on the wet side of the instrument
Localize the leakage and take care of the leakage, for example
by tightening connectors.
Air in pumps
Purge pumps. See Section 5.4.1 System pumps, on page 162.
Bad pump function
See troubleshooting of pumps in Section 8.6.
Leakage inside the Frac chamber
Contact Service.
The fraction volume found
in the tubes or wells are
smaller than expected
The fingertight connector on the Dispenser head is not tight
enough
Tighten the connector. Replace the connector if the leakage is
persistent.
The waste tubing or the waste funnel is blocked.
Make sure that there is no blockage or clogging and that the
waste tubing not is bent.
Leakage in the Frac arm
Contact service.
Liquid on the floor of the
fraction collector
382 ÄKTA pure User Manual 29119969 AB
8 Troubleshooting
8.5 Troubleshooting: Fraction collector