User Manual Druck DPI620 Genii
Figure 2-23 CH1 Thermocouple Source CH2 Switch Test
Figure 2-23 shows a thermocouple switch test.
1. Set the applicable software options:
The TC is set to source a temperature
The UTILITY is set to SWITCH TEST. The AUTOMATION is
set to RAMP
2. Complete the electrical connections.
3. It is a CH1 function, the switch connections must be on
CH2.
4. For the Ramp process, set START and STOP values that
are applicable to the switch value.
5. To get an accurate switch value, set a long TRAVEL
period.
06 June 2013 K0541 issue 1 Page 58 of 175

Table of Contents