5-58 Issue 01, 04/2005 Krautkramer USM 35X
Operation
HORN
In this function, you can decide whether or not an
acoustic alarm should be given in addition to the visual
alarm (LED A).
– Select the function HORN.
– Use the left-hand rotary knob to set the horn to on or
off.
H Note:
Error alarms can be triggered unter certain circum-
stances. These are caused by intermediate conditions
in instrument operation occuring when the instrument is
used, i.e. when function parameters are changed. Pos-
sible alarms occuring during instrument operation
(setting of functions) are to be ignored.
General configuration
EVAMOD (Echo evaluation)
This is where you can choose a method for the evalua-
tion of the measured reflector echo. Depending on the
instrument version used, you have various methods to
choose from.
• REF (default setting)
Evaluation using the measurement of dB difference,
available for all instrument versions
• AWS
Rating of welds according to AWS
• DAC (only USM 35X DAC and USM 35X S)
Evaluation using the Distance-Amplitude Curve
• JISDAC (only USM 35X DAC and USM 35X S)
Evaluation using the Distance-Amplitude Curve
according to JIS Z3060-2002
• DGS (only USM 35X S)
Evaluation using the DGS method
– Select the function EVAMOD.
– Use the right-hand rotary knob to choose the required
method.
5-58 Issue 01, 04/2005 Krautkramer USM 35X
Operation
HORN
In this function, you can decide whether or not an
acoustic alarm should be given in addition to the visual
alarm (LED A).
– Select the function HORN.
– Use the left-hand rotary knob to set the horn to on or
off.
H Note:
Error alarms can be triggered unter certain circum-
stances. These are caused by intermediate conditions
in instrument operation occuring when the instrument is
used, i.e. when function parameters are changed. Pos-
sible alarms occuring during instrument operation
(setting of functions) are to be ignored.
General configuration
EVAMOD (Echo evaluation)
This is where you can choose a method for the evalua-
tion of the measured reflector echo. Depending on the
instrument version used, you have various methods to
choose from.
• REF (default setting)
Evaluation using the measurement of dB difference,
available for all instrument versions
• AWS
Rating of welds according to AWS
• DAC (only USM 35X DAC and USM 35X S)
Evaluation using the Distance-Amplitude Curve
• JISDAC (only USM 35X DAC and USM 35X S)
Evaluation using the Distance-Amplitude Curve
according to JIS Z3060-2002
• DGS (only USM 35X S)
Evaluation using the DGS method
– Select the function EVAMOD.
– Use the right-hand rotary knob to choose the required
method.