5-18 Issue 01, 04/2005 Krautkramer USM 35X
Operation
Calibrating the USM 35X
Calibration with straight- and
angle-beam probes
Case A: With known material velocity
Calibration process
Set the known material velocity in MTLVEL (function
group BASE).
Couple the probe to the calibration block.
Set the required display range in RANGE (function
group BASE). The calibration echo must be dis-
played on the screen.
Position the gate on one of the calibration echoes
until the sound path of the echo is indicated in the
measurement line.
After this, change the adjustment of the function
P-DELAY (function group BASE) until the correct
sound path for the selected calibration echo is
indicated in the measurement line.
Example:
You are carrying out the calibration for the calibration
range of 100 mm/5" via the function group BASE using
the calibration block V1 (thickness 25 mm/1") which is
laid flatwise.
Set RANGE to 100 mm/5".
Set the known material velocity of 5920 m/s
(233 "/ms) in MTLVEL.
Set the gate so that it is positioned on the first
calibration echo (from 25 mm/1").
Read the sound path in the measurement line. If this
value is not equal to 25 mm/1", change the adjust-
ment for the function P-DELAY until it is at 25 mm/1".
This completes the calibration of the USM 35X to the
material velocity of 5920 m/s (233 "/ms) with a calibra-
tion range of 100 mm/5" for the probe used.
5-18 Issue 01, 04/2005 Krautkramer USM 35X
Operation
Calibrating the USM 35X
Calibration with straight- and
angle-beam probes
Case A: With known material velocity
Calibration process
Set the known material velocity in MTLVEL (function
group BASE).
Couple the probe to the calibration block.
Set the required display range in RANGE (function
group BASE). The calibration echo must be dis-
played on the screen.
Position the gate on one of the calibration echoes
until the sound path of the echo is indicated in the
measurement line.
After this, change the adjustment of the function
P-DELAY (function group BASE) until the correct
sound path for the selected calibration echo is
indicated in the measurement line.
Example:
You are carrying out the calibration for the calibration
range of 100 mm/5" via the function group BASE using
the calibration block V1 (thickness 25 mm/1") which is
laid flatwise.
Set RANGE to 100 mm/5".
Set the known material velocity of 5920 m/s
(233 "/ms) in MTLVEL.
Set the gate so that it is positioned on the first
calibration echo (from 25 mm/1").
Read the sound path in the measurement line. If this
value is not equal to 25 mm/1", change the adjust-
ment for the function P-DELAY until it is at 25 mm/1".
This completes the calibration of the USM 35X to the
material velocity of 5920 m/s (233 "/ms) with a calibra-
tion range of 100 mm/5" for the probe used.