Krautkramer USM 35X Issue 01, 04/2005 5-9
Operation
5.4 Adjusting the pulser
(function group PULS)
You will find all functions for the adjustment of the puls-
er in the function group PULS.
– If required, go to the first operating level.
– Select the function group PULS.
DAMPING (Probe matching)
This function serves for matching the probe. You can
use it to adjust the damping of the probe’s oscillating
circuit and to consequently change the height, width
and resolution of the echo display.
• low
This setting has a lower damping effect and produc-
es higher and broader echoes.
• high
This setting reduces the echo height but mostly also
produces narrow echoes with higher resolution.
– Select the function DAMPING.
– Set the required value by means of the right-hand
rotary knob.
Adjusting the pulser (function group PULS)
Krautkramer USM 35X Issue 01, 04/2005 5-9
Operation
5.4 Adjusting the pulser
(function group PULS)
You will find all functions for the adjustment of the puls-
er in the function group PULS.
– If required, go to the first operating level.
– Select the function group PULS.
DAMPING (Probe matching)
This function serves for matching the probe. You can
use it to adjust the damping of the probe’s oscillating
circuit and to consequently change the height, width
and resolution of the echo display.
• low
This setting has a lower damping effect and produc-
es higher and broader echoes.
• high
This setting reduces the echo height but mostly also
produces narrow echoes with higher resolution.
– Select the function DAMPING.
– Set the required value by means of the right-hand
rotary knob.
Adjusting the pulser (function group PULS)