5-68 Issue 01, 04/2005 Krautkramer USM 35X
Operation DAC according to JIS Z3060-2002 (only USM 35X DAC and USM 35S)
DACECHO (Recording reference curve)
A Attention:
Before starting to record a reference curve, the instru-
ment must be correctly calibrated (ref. section 5.7
Calibrating the USM 35X
).
The moment a new curve is recorded, a possibly al-
ready existing curve must be deleted. If necessary,
make sure that the old curve has been stored in a free
data set before starting to record a new curve!
Select the function DACMOD.
Use the right-hand rotary knob to set the function to
DAC. The function DACECHO is set to 0 since there
is no previously recorded echo.
Couple the probe to the reference block, and peak
the first reference echo. Use the left-hand rotary knob
to bring the echo to an amplitude between 70 % and
100 % screen height.
Select the function aSTART, and then move the gate
so that the selected echo is the highest of the echo
sequence within the gate range.
Press to record the first reference echo. The
instrument gain will automatically change until the
DAC echo in gate A reaches 80% screen height
(+/– 0,3 dB). The function DACECHO is set to 1 to
indicate that the first reference echo has been
successfully recorded. Simultaneously the status
symbol “R” appears (= reference echo stored).
H Note:
The dB-value by which the gain has been changed re-
lated to the reference gain can be displayed using the
new parameter DAC dB.
5-68 Issue 01, 04/2005 Krautkramer USM 35X
Operation DAC according to JIS Z3060-2002 (only USM 35X DAC and USM 35S)
DACECHO (Recording reference curve)
A Attention:
Before starting to record a reference curve, the instru-
ment must be correctly calibrated (ref. section 5.7
Calibrating the USM 35X
).
The moment a new curve is recorded, a possibly al-
ready existing curve must be deleted. If necessary,
make sure that the old curve has been stored in a free
data set before starting to record a new curve!
Select the function DACMOD.
Use the right-hand rotary knob to set the function to
DAC. The function DACECHO is set to 0 since there
is no previously recorded echo.
Couple the probe to the reference block, and peak
the first reference echo. Use the left-hand rotary knob
to bring the echo to an amplitude between 70 % and
100 % screen height.
Select the function aSTART, and then move the gate
so that the selected echo is the highest of the echo
sequence within the gate range.
Press to record the first reference echo. The
instrument gain will automatically change until the
DAC echo in gate A reaches 80% screen height
(+/– 0,3 dB). The function DACECHO is set to 1 to
indicate that the first reference echo has been
successfully recorded. Simultaneously the status
symbol “R” appears (= reference echo stored).
H Note:
The dB-value by which the gain has been changed re-
lated to the reference gain can be displayed using the
new parameter DAC dB.