5-74 Issue 01, 04/2005 Krautkramer USM 35X
Operation
Define your settings:
PROBE-#: probe number
Fixed-programmed probes with all settings (PRB-
NAME, DEL-VEL, D eff and PRBFREQ cannot be
changed in the case of these probes);
PROBE-# = 0 is user-programmable with reference
to all parameters.
PRBNAME: probe name
The name belongs to the selected probe number and
cannot be changed; the individual probe name can be
entered only with PROBE-# = 0.
DGS-CRV: registration curve for DGS evaluation
This enables you to select the circular disk-shaped
equivalent reflector diameter to be used for displaying
the DGS curve and used as recording threshold for
echo evaluations.
DEL-VEL: delay material velocity of the probe
Predefined with programmed probes.
D eff: effective element diameter of the probe used
Predefined with programmed probes.
PRBFREQ: probe frequency
Frequency of the transducer; predefined with pro-
grammed probes.
REFECHO: type of the reference reflector used
BW backwall
SDH side drilled hole
FBH flat bottom hole
REFSIZE: size of the reference reflector
ATT-REF: sound attenuation in the reference block
ATT-OBJ: sound attenuation in the test object
AMPLCOR: value for the amplitude correction.
This is required whenever you are using an angle
probe and the quadrant echo from the calibration
standards K1 or K2 as a reference reflector.
To return to the A-scan, press one of the following
keys:
, or .
Evaluation according to the DGS method (only USM 35X S)
5-74 Issue 01, 04/2005 Krautkramer USM 35X
Operation
Define your settings:
PROBE-#: probe number
Fixed-programmed probes with all settings (PRB-
NAME, DEL-VEL, D eff and PRBFREQ cannot be
changed in the case of these probes);
PROBE-# = 0 is user-programmable with reference
to all parameters.
PRBNAME: probe name
The name belongs to the selected probe number and
cannot be changed; the individual probe name can be
entered only with PROBE-# = 0.
DGS-CRV: registration curve for DGS evaluation
This enables you to select the circular disk-shaped
equivalent reflector diameter to be used for displaying
the DGS curve and used as recording threshold for
echo evaluations.
DEL-VEL: delay material velocity of the probe
Predefined with programmed probes.
D eff: effective element diameter of the probe used
Predefined with programmed probes.
PRBFREQ: probe frequency
Frequency of the transducer; predefined with pro-
grammed probes.
REFECHO: type of the reference reflector used
BW backwall
SDH side drilled hole
FBH flat bottom hole
REFSIZE: size of the reference reflector
ATT-REF: sound attenuation in the reference block
ATT-OBJ: sound attenuation in the test object
AMPLCOR: value for the amplitude correction.
This is required whenever you are using an angle
probe and the quadrant echo from the calibration
standards K1 or K2 as a reference reflector.
To return to the A-scan, press one of the following
keys:
, or .
Evaluation according to the DGS method (only USM 35X S)