Krautkramer USM 35X Issue 01, 04/2005 5-77
Operation
S-DISP = Ha%, i.e. echo height evaluation as a per-
centage.
At the same time, the multiple-curve display mode is
also active with a curve distance of OFFSET = 6 dB.
S-DISP = Ha%Crv, the evaluation result is now the
value exceeding the curve in %.
S-DISP = Gt is a special case: the value displayed is
the DGS test sensitivity with which the maximum of the
specified curve is at 80 %.
This value serves for the purpose of checking and doc-
umenting.
Evaluation according to the DGS method (only USM 35X S)
Krautkramer USM 35X Issue 01, 04/2005 5-77
Operation
S-DISP = Ha%, i.e. echo height evaluation as a per-
centage.
At the same time, the multiple-curve display mode is
also active with a curve distance of OFFSET = 6 dB.
S-DISP = Ha%Crv, the evaluation result is now the
value exceeding the curve in %.
S-DISP = Gt is a special case: the value displayed is
the DGS test sensitivity with which the maximum of the
specified curve is at 80 %.
This value serves for the purpose of checking and doc-
umenting.
Evaluation according to the DGS method (only USM 35X S)