7 - 6 Section 7-4 - Screen Captures
7-4-1 Check and Record the P4 Key Function (cont’d)
5.) In the Destinations section, record the service that is displayed. The destinations list displays
the following information:
* Name: user defined during service configuration
* Type: the type of service
* Server: the device for which the service was configured
* Dir: direction: output, input, or both (I+O)
NOTE: From BTO5 software, Check and Record the P4 Key Function
In the Physical Print Buttons section, record the service that is displayed. The Physical Print Buttons
section displays the following information :
* Format : the type of image format
* Image Frames : Single, Multiple, Secondary Capture
* Capture Area : Video Area, Image Area, Whole Screen
* Compression : The type of image compression.
* Quality % : Image Quality.
In the Physical Print Buttons section, record the parameters related to the service.
6.) In the Image generated section, record the parameters related to the service.
NOTE: From BTO5 software,In the Physical Print Buttons section, record the parameters related to the
service.
7-4-2 Setting the P4 Key to Screen Capture
If the P4 Key is not set to screen capture:
1.) Go to Utility->Connectivity->Button.
2.) Select Print4 under Physical Print Buttons section and set the Image Frames to Secondary
Capture. Add “Copy to Dataflow” to printflow view.
3.) Ensure that the Image Frames is set to Secondary Capture, Capture Area to Whole Screen,
and Compression to None.
4.) The P4 Key should now be set up for whole screen capture, sending the screens to the image
buffer (clipboard).