GE MEDICAL SYSTEMS
D
IRECTION 2380207, REVISION 7 LOGIQ™ 5 PRO SERVICE MANUAL
Chapter 8 Replacement Procedures 8 - 49
8-5-5 Top Cover (FRU No. 304)
Purpose: This is a description on how to remove and replace the Top Cover.
8-5-5-1 Tools
• Common pillips screwdrivers
8-5-5-2 Needed Manpower
• 1 person, 20 minutes + travel
8-5-5-3 Preparations
• Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
8-5-5-4 Removal Procedure
1.) Remove the Left Cover (FRU No.300). Refer to section 8-5-1 on page 41.
2.) Remove the Right Cover (FRU No.301). Refer to section 8-5-2 on page 43.
3.) Remove the Rear Cover (FRU No.302). Refer to section 8-5-3 on page 45.
4.) Unscrew two screws (1-2) from the top side of the Top cover. Refer to Figure 8-47.
Figure 8-47 Unscrew 2 screws from the Top cover
(1)
(2)