IRECTION 2380207, REVISION 7 LOGIQ™ 5 PRO SERVICE MANUAL
Chapter 8 Replacement Procedures 8 - 27
8-4-6 OP Panel TGC Assy (FRU No. 205)
Purpose: This is a description on how to remove and replace OP Panel TGC Assy.
8-4-6-1 Tools
• Common pillips screwdrivers.
8-4-6-2 Needed Manpower
• 1persons, 5 minutes + travel
8-4-6-3 Preparations
• Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
• Keyboard Assy should be removed before proceed. For more information, refer to section 8-4-1 on
page 16.
8-4-6-4 Removal Procedure
1.) Remove eight (8) TGC knobs from the upper side of the Keyboard. Refer to Figure 8-27.
2.) Disconnect the TGC connector from the back side of the keyboard.
3.) Unscrew four (4) screws and Remove the TGC Assy from the keyboard.
Figure 8-27 Removing TGC Knobs
Figure 8-28 Unscrew 4 screws and remove TGC Assy
TGC Connector
1
2
3
4