GE MEDICAL SYSTEMS
DIRECTION 2380207, REVISION 7 LOGIQ™ 5 PRO SERVICE MANUAL
8 - 26 Section 8-4 - Keyboard Block
8-4-5-4 Removal Procedure (cont’d)
5.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-4-5-5 Mounting Procedure
Install the new parts in the reverse order of removal.
Table 8-12 Functional Tests
Service Manual
Section Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
2380207, Rev 7+, Section 8-4-5. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-11-2
OP Panel Encoder Assy Validation