EasyManua.ls Logo

GE LOGIQ 5 PRO - Section 8-10 Software Loading Procedure

GE LOGIQ 5 PRO
462 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
DIRECTION 2380207, REVISION 7 LOGIQ™ 5 PRO SERVICE MANUAL
8 - 166 Section 8-10 - Software Loading Procedure
Section 8-10
Software Loading Procedure
8-10-1 General
This describes a full system software loading (Base Software Load Image + LOGIQ 5 PRO Application)
procedure.
Use this instruction when:
- Performing the HDD with new one (Base Software Load Image + Application)
- Performing the system revision-up (Application)
Reference :
The below is the support table between Application Software version and Base System
Software Load Image part number
8-10-2 Parts Required
Base System Software Load Image CD
LOGIQ 5 PRO R4.x.x Application Software
Blank CD-R ( for patient image backup)
Service Dongle
NOTE: Patient Image Data will be deleted when all the drives are ghosted. Back up them on the CD-R
or MOD before starting software loading
NOTE: Insert the service dongle into ths service port located at the rear panel before installing the Base
System Software Load image
8-10-3 TimeRequired
Approximately 2 hour
Table 8-55 Base Software Load Image CD Part Numbers
Application
Software Version
Base Software Load Image Part P/N
5134559-x (BEP2)
5132191-x (BEP3)
R4.X.X
Supported Supported

Table of Contents

Related product manuals