IRECTION 2380207, REVISION 7 LOGIQ™ 5 PRO SERVICE MANUAL
Chapter 4 Functional Checks 4 - 3
4-3-1-1 Scanner Power On (cont’d)
NOTE: When power is applied to the Scanner, and the Rear Circuit breaker is turned ON, Power is
distributed to the ATX smps in the Back End Processor and Power control circuit in the AC
power assy . When the Power ON/OFF key is pressed once, the power control circuit turn on
the card cage and fans, and Back End Processor ,Monitor. Backend Processor starts and its
software code is distributed to initiate the scanner.
Figure 4-24 Circuit Breaker
Main Power Cable
Main Circuit Breaker