GE MEDICAL SYSTEMS
D
IRECTION 2380207, REVISION 7 LOGIQ™ 5 PRO SERVICE MANUAL
Chapter 8 Replacement Procedures 8 - 85
8-6-5 Bumper Set (FRU No. 405)
Purpose: This is a description on how to remove and replace the Bumper Set.
8-6-5-1 Tools
• Common pillips screwdrivers
8-6-5-2 Needed Manpower
• 1persons, 20 minutes + travel
8-6-5-3 Preparations
• Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
8-6-5-4 Removal Procedure
1.) Remove the Monitor Assy (FRU No.100). Refer to section 8-3-1 on page 2.
2.) Remove the Left Cover (FRU No.300). Refer to section 8-5-1 on page 41.
3.) Remove the Right Cover (FRU No.301). Refer to section 8-5-2 on page 43.
4.) Remove the Front Base Cover (FRU No.305). Refer to section 8-5-6 on page 51.
5.) Unscrew two screws (1-2).
6.) Remove the Bumper Set. Refer to Figure 8-77.
Figure 8-77 Removing the Bumper Set
1
2