GE MEDICAL SYSTEMS
DIRECTION 2380207, REVISION 7 LOGIQ™ 5 PRO SERVICE MANUAL
8 - 50 Section 8-5 - Covers
8-5-5-4 Removal Procedure (cont’d)
5.) Unscrew four screws (3-6) from the bottom of the Top cover. Refer to Figure 8-48.
6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-5-5-5 Mounting procedure
Install the new parts in the reverse order of removal.
Figure 8-48 Unscrew 4 screws from the bottom
Table 8-24 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
2380207, Rev 7+, Section 8-5-5. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-12-1
Cover Parts Function Validation
Section 10-5-5
Physical Inspection
(3)
(4)
(5)
(6)