GE RAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
3-90 Section 3-13 - Network Configuration
2.) Select the Configuration page, then double-click InSite Exc Agent Configuration(A).
NOTE: This example shows fictional numbers!
3.) Fill out at least all bold stated mandatory fields in the Agent Configuration section(B).
- The Device Name9Pre-populated) and the CRM field have to be prefix KP.
- Select Continent and Country from the pull-down menu.
- Enter City, State, Postal Code and Institution where the system resides.
4.) Check settings in the Advanced Configuration section (C).
- Enterprise Server - PRODUCT
- Service Center - EURO
- Log Level - WARN
Further fields should be pre-populated and should not be modified. However, please ensure correct
setting.
- Enterprise Server URL - pre-populated URL.
- Enterprise Tunnel URL -pre-populated URL.
- File Repository - This path is set by engineering and must not be changed!
- Dir-D:\InsiteUploadFolder
- Filter-ensure that this field has the entry “*.txt”s
- File Watcher: Should always be Enabled!
Figure 3-91 Common Service Desktop - Configuration