GE RAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
8-180 Section 8-57 - Replacement of ECG Kit assy
Section 8-57
Replacement of ECG Kit assy
8-57-1 Manpower
One person, 10 minutes.
8-57-2 Tools
Philips screwdriver
8-57-3 Preparations
Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
8-57-4 Removal Procedure
1.) Remove 5 probe holders.
2.) Unscrew 9 numbers of screws below the keyboard base.
Figure 8-254 Removing 5 Probe holders and unscrewing 9 screws below the keyboard base