GE RAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
8-24 Section 8-7 - Replacement or Activation of Options
8-7-1 Operation for activating Options
1.) Press the F2 key on the control panel.
2.) In the “Utilities” menu select SYSTEM SETUP
to invoke the setup desktop on the screen.
3.) On the left side of the screen select ADMINISTRATION
and then click the OPTIONS tab (see:
Figure 8-2 on page 8-7.)
8-7-1-1 Operation for installing a “Demo Key” or a “Permanent Key”:
1.) Position the cursor inside the input field desired and press the right/left trackball
key.
2.) If one exists, clear/edit the current key code.
3.) Enter the encrypted serial code with the keyboard and click on SUBMIT
. (Code will be checked.)
4.) Click the SAVE&EXIT
button.
NOTE: After activating a key code, restart (turn off and on) the Voluson™ P8 / Voluson™ P6 system.